Anesthesia System

Anesthesia System

0 out of 5

Application:

The Anesthesia machine makes a good performance in Intensive Care Units (ICU), Operation room, Anesthesiology Department and other departments. From high to low acuity, simple to complex cases, pediatric to geriatric patients, Superstar Medical Anesthesia Systems offer you the choices you need in ventilation, monitoring and technique.

Description

Features

  • l 10.4” TFT LCD screen displays the Ventilation parameters, Alarming information and Waveform.
  • l High precision flow meter, instantly know the fresh gas flow to your patient.
  • l Integration breathing circuit design, ensure easy operating and keep tidy.
  • l Multiple working modes such as volume control and pressure limit, adapt to wide range patient.
  • l Vaporizer with temperature, flow compensation and self-lock function, keep safety anytime.
  • l Multiple parameters monitoring interface, make every parameter clear, let users know the patient conditions in all aspects;
  • l Real time pressure-time, flow-time loop graphics and high precision Oconcentration detection function included.
  • l Vital sign monitor and Anesthetic gas monitor are optional.
  • l ETCOmodule is optional.

Safety

  • l Three level alarming system, visual and sound alarm information.
  • l With lots of alarming, reminding and protection functions.
  • l Advanced power management control technology.
  • l With built-in backup power source, when outside power source goes off, back-up power source starts to work.
  • l Low O2 pressure alarm and N2O cut-off protection.

Specifications

Ventilation modes: IPPV; A/C; PCV; SIMV; SIGH; MAUN
Ventilator parameter ranges
Flow meter O2(0.1-10L/min)
N2O(0.1-10L/min)
AIR(0.1-10L/min)
Rapid oxygen supply 35L/min-75L/min
Tidal volume(Vt) 0, 20 mL ~ 1500 mL
Frequence(Freq) 1bpm ~ 100 bpm
I/E 4:1~1:8
PEEP 0cmH2O ~ 30 cmH2O
Pressure triggering sensitivity(PTr) -20 cmH2O ~ 0 cmH2O(Based on PEEP)
Flow trigger sensitivity (FTr) 0.5 L/min ~ 30 L/min
Pressure control (PC) 5 cmH2O ~ 60 cmH2O
SIGH 0(off)  1/100 ~ 5/100
Apnea Ventilation OFF,5 s ~ 60 s
Pressure Limit 20 cmH2O ~ 100 cmH2O
Monitored Parameters
Frequence(Freq) 0 /min ~ 100 /min
Tidal volume(Vt) 0 mL ~ 2000 mL
MV 0 L/min ~ 100 L/min
Oxygen concentration 15 % ~ 100 %
Graphical display:
P-T(pressure – time)
F-T(flow – time)
P-V loop (pressure – volume loop)
Size

Wooden case packing size:L 770*W 890*H 1535mm , GW:151KG ; NW:91.2KG

Alarm and protection

The AC power failure alarm Power failure or no connection
Internal battery backup low voltage alarm <11.3±0.3V
No tidal volume ≤5Ml within 6s
High oxygen concentration alarm

Low oxygen concentration alarm

19%-100%

18%-99%

High Airway  pressure alarm

Low Airway  pressure alarm

High Minute Volume alarm

Low Minute Volume alarm

Continuous Pressure alarm

20cmH2O-100cmH2O

0cmH2O-20cmH2O

Adult(5L/min-20L/min) Paed(1L/min-15L/min

0-10L/min)

(PEEP+1.5kPa) over 16s

Suffocation warning 5s-60s no spontaneous ventilation
The maximum limited pressure <12.5 kPa
Fan error

Oxygen deficit

Show on screen

Show on screen

Working conditions
Gas source O2,N2O,AIR
Pressure 280kPa-600kPa
Voltage 100-240V
Power frequency 50/60Hz
Input power 85VA
Vaporizer